Amerikanische Gesellschaft für Hirudotherapie

Near infrared spectroscopy in monitoring of head and neck microvascular free flaps

Research article published in Bratislavske lekarske listy (2023)

Zuletzt aktualisiert: June 18, 2026Geprüft von: ASH Editorial Board
Research article — evidence reviewArticle reference
Evidence: Research reportArzneimittelentwicklungCzako L et al. · Bratislavske lekarske listy, 2023

Abstract

OBJECTIVES: Microvascular free‑flap monitoring is crucial to the early detection of flap failure and increases the chance of early intervention in case of disruption of perfusion to a flap. Many clinical alternatives to classical clinical flap monitoring have been proposed, such as color duplex ultrasonography, handheld Doppler, flap thermometry, or implantable Doppler flowmetry. Early detection of critical changes in tissue oxygenation can lead to successful surgical intervention when problems with flap nutrition arise. METHODS: Our clinical study seeks to investigate dynamic monitoring of free flaps with near‑infrared spectroscopy (NIRS). NIRS is a non-invasive instrumental technique used for continuous monitoring of peripheral tissue oxygenation (StO2) and microcirculation. All patients were included prospectively from one clinical center. RESULTS: During the clinical research period, 18 patients underwent extraoral head and neck reconstruction with one of three types of free flap, namely with radial forearm free flap (RFFF), anterolateral thigh flap (ALT) or fibula free flap (FFF). Measurements of flap perfusion were taken using NIRS during intraoperative and postoperative phases for 71 hours on average. A total of 6 perfusion disorders were recorded, of which three originated from microanastomoses and three from postoperative bleeding and compression of pedicle. NIRS showed characteristic changes in all 6 cases that were returned to the operating theatre owing to pedicle compromise. In these cases, NIRS had detected the pedicle compromise before it was clinically identified. A single StO2 monitoring was able to detect the vascular compromise with 100% sensitivity and 95.65% specificity. None of the cases were falsely positive. In our study, all compromised flaps were accurately identified by means of NIRS. In most cases, the changes in oxygen saturation became evident on NIRS prior to being clinically observed. CONCLUSION: In our study, the continuous NIRS monitoring securely detected the early stages of arterial and venous thromboses or pedicle compression. The most important aspects of monitoring the flaps´ microvascular perfusion and vitality by means of NIRS lie in its function of recording the dynamics of changes in the values ​​of absolute oxygen saturation (StO2> 50%) alongside with detecting a 30% decrease in tissue saturation over a 60‑minute interval (60 min StO2 >30%) before the clinical changes in the microvascular flap become observable. In cases of pedicle compression, the average time of appearance of signs of StO2 values dropping below the reference interval (as detected by NIRS) was 1:29:02 hour (SD= 0:58:42 h) prior to the occurrence of any clinical signs, while in cases of microvascular anastomosis complications, it was 0:35:23 hour (SD=0:08:30 h)  (SD = 0:08:30 h) (Tab. 3, Fig. 7, Ref. 42).

Abstract sourced from PubMed (NCBI) for the cited record. See the original publication for the authoritative version.

Publication typeJournal Article
Indexed MeSH termsHumansFree Tissue FlapsSpectroscopy, Near-InfraredPlastic Surgery ProceduresPostoperative Complications

Zusammenfassung

Microvascular freeflap monitoring is crucial to the early detection of flap failure and increases the chance of early intervention in case of disruption of perfusion to a flap.

Warum dies für die Hirudotherapie relevant ist

This prospective clinical study evaluated NIRS monitoring of tissue oxygen saturation (StO2) in 18 patients undergoing head and neck reconstruction with radial forearm, anterolateral thigh, or fibula free flaps, recording 6 perfusion disorders over an average 71-hour monitoring period. NIRS detected all cases of pedicle compromise with 100% sensitivity and 95.65% specificity, identifying changes approximately 1.5 hours before clinical signs in compression cases and 35 minutes in microanastomosis cases. For ASH's domain, the relevance is indirect—early detection of free-flap vascular compromise could prompt interventions including leech therapy for venous congestion, but the abstract contains no mention of leeches or hirudotherapy. The caveat is that this study focuses on optical monitoring technology for flap surveillance with no leech involvement.

Zitation

Near infrared spectroscopy in monitoring of head and neck microvascular free flaps

Czako L et al. · Bratislavske lekarske listy, 2023

Verwandter klinischer Kontext

Erfahren Sie, wie diese Forschung mit der klinischen Praxis verknüpft ist

Zur ASH-Bibliothek hinzugefügt: May 27, 2026 · Letzte Aktualisierung der Website: June 18, 2026

Diese Website stellt Bildungsinformationen bereit und ist weder eine medizinische Beratung noch eine Diagnose oder Behandlungsempfehlung. Die medizinische Blutegeltherapie ist mit klinisch relevanten Risiken verbunden und sollte ausschließlich von qualifizierten Klinikerinnen und Klinikern unter institutionell genehmigten Protokollen durchgeführt werden. Die FDA-510(k)-Zulassung für medizinische Blutegel ist auf bestimmte Indikationen beschränkt; experimentelle und Off-Label-Diskussionen werden entsprechend gekennzeichnet. Für patientenspezifische Beratung wenden Sie sich an eine qualifizierte Gesundheitsfachkraft.