Thrombosis during off pump LVAD placement in a patient with heparin induced thrombocytopenia using bivalirudin
Case report published in Journal of cardiothoracic surgery (2013)
Abstract
Here we present our attempt at off pump HeartMate II left ventricular assist device (LVAD) implantation using the anticoagulant bivalirudin in a patient with heparin induced thrombocytopenia, which resulted in thrombosis within the LVAD device. This required that our procedure be converted to on pump, and a new HeartMate II LVAD device to be implanted. In our view, this thrombotic event may have been caused by a number of factors that include bivalirudin's (1) short half-life of about 20 minutes, (2) decreased activity with blood stasis, (3) inability to prevent clot propagation, and (4) uncertainty with real-time monitoring of therapeutic levels. To prevent future thrombotic events, it may be beneficial to immediately deair the LVAD device prior to the coring of the left ventricle during off pump LVAD placement. In addition, a solution other than blood may be used for priming. If blood is used for priming of the LVAD device, the duration of blood stasis should not exceed 20 minutes when bivalirudin is being used for anticoagulation. Furthermore, this case emphasizes the importance of having a backup LVAD device available and ready to use during surgical procedures.
Abstract sourced from PubMed (NCBI) for the cited record. See the original publication for the authoritative version.
Резюме
Here we present our attempt at off pump HeartMate II left ventricular assist device (LVAD) implantation using the anticoagulant bivalirudin in a patient with heparin induced thrombocytopenia, which resulted in thrombosis within the LVAD device.
Почему это важно для гирудотерапии
This case report describes a patient with heparin-induced thrombocytopenia who underwent off-pump HeartMate II LVAD implantation using bivalirudin for anticoagulation, resulting in intra-device thrombosis that required conversion to on-pump surgery and device replacement. The authors attribute the failure to bivalirudin's short half-life (~20 minutes), decreased activity during blood stasis, limited ability to prevent clot propagation, and monitoring challenges, and offer procedural recommendations for future surgeries. The abstract does not mention hirudin, leeches, leech therapy, or the leech secretome; it discusses bivalirudin solely as an anticoagulant used in cardiac surgery. There is no defensible connection to hirudotherapy or ASH's domain based on the abstract alone.
Цитирование
Thrombosis during off pump LVAD placement in a patient with heparin induced thrombocytopenia using bivalirudin
Awad H et al. · Journal of cardiothoracic surgery, 2013
Связанный клинический контекст
Узнайте, как это исследование связано с клинической практикой
Добавлено в библиотеку ASH: May 27, 2026 · Последнее обновление сайта: June 18, 2026